Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the new-album-gallery domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/u725103605/domains/wedigitalcreatives.com/public_html/vas/wp-includes/functions.php on line 6121

Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the metform domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/u725103605/domains/wedigitalcreatives.com/public_html/vas/wp-includes/functions.php on line 6121
Platinum – VAS Spectrometers

Platinum - Overview

The Platinum Series comes with the following salient capabities.

Performance. Period. The Quanta-Platinum Series of spectrometers/Metal Analyzers have been designed from the ground up to meet the most demanding metal testing requirements in the Industry. With N2 and O2* analysis, the Quanta-Platinum Series is the world’s most economic ultra-high performance spectrometer on the market today.

Product Video

Technical

Excitation Source

– V-MAP current controlled source with true 100KHz real-time regulation system
– Fully software-controlled excitation source
– Multi-frequency spark source, up to 1KHz
– High energy micro-melt for better precision (internally software controlled)
– High precision spark condensed spark source
– Analytical program parameters selected automatically by the software
– Maintenance free

Software Analysis

– Extremely user-friendly operation
– Factory calibrated
– Automatic inter-element interface corrections
– Display of Mean SD or RSD
– User flexibility to standardize each analytical program
– User configurable type standardization
– Logging of the analysis data
– Quick retrieval of results from historical data
– User-friendly report generation
– Option to transfer data to Excel spreadsheet

Optical System

– Rowland circle arrangement 
– Focal Length: 400mm or 500mm
– Resolution: > 3 pm 
– Wavelength range: 126nm to 800nm
– Temperature stabilized and sealed against dust and contamination
– High-performance holographic diffraction grating
– High-resolution multiple CCD detector (up to 32 CCDs) arrays with 3694pixels each
– Profiling – automatic & real-time, software controlled

Spark Stand

– Specially designed C-shaped sample plate to prevent warping and deformations over long periods of usage.
– Argon flushed with automatically adjusted low flow in standby and analytical modes
– Tungsten electrode
– Mechanical quick-engage sample clamp – Flexible for large and small sample sizes
– Easily removable sample base for chamber cleaning and maintenance

Control & Data Processing

– Ethernet and USB connectivity.
– Windows® based software. Works with Win 7 and above.
– Complete and accurate control of instrument through a high performance 32 bit Atmel ATxMega-based data acquisition system
– Works with standard desktop personal computer (connected externally)

Weight & Dimensions

– Instrument Weight 80kg approx. (without utility base)
– Instrument size 1000 X 700 x 380 (W x Dx H)

Environmental Requirement

– Operating temperature 22 to 28°C
– Storage temperature -10 to 70°C

Electrical Requirement

– 1 phase 230V AC, 50Hz, 6 Amps (available in all international power Specifications)